4.1 Measurement of Electrical Biosignals
|
97
Electrode
Insolation
Outer electrode
Inner elctrode
Fig. 4.5: Schematic representation of
monopolar (left) and bipolar (right) elec-
trode needle (modified from [50]).
Tab. 4.2: Signal amplitude and bandwidth of selected electrical biosignals.
Biosignal
Electrode type
Signal amplitude
Bandwidth in Hz
Electrocardiogram (ECG)
Skin electrode
0,1–4 mV
0,01–250
Electroencephalogram (EEG
Skin electrode
1–100 µV
0,01–100
Electromyogram (EMG)
Needle electrode
0,1–5 µV
1–5.000
Electrooculogram (EOG)
Skin electrode
0,05–4 mV
0,01–100
Electroneurogram (ENG)
Needle electrode
0,01–3 mV
0–10.000
4.1.2 Electrical Amplifier
Depending on the biosignal and the electrode position,thedifferentialvoltagebetween
the measuring electrodes at the body surface ranges from a few microvolts for the EEG
to several millivolts for the R-wave in the ECG. The bandwidth of electrical biosignals
ranges from 0 Hz to a maximum of 10 kHz. In Table 4.2 values for signal amplitude
and bandwidth of some important biosignals are listed.
The amplifier has the task of raising the biosignal to a voltage level in the volt
range with a constant gain over the entire bandwidth. As a rule, the biosignal (useful
signal) is superimposed by interference signals such as the common-mode signal or
noise. A common-mode interference is caused, for example, by power lines located
in the vicinity of the measurement setup. The origin of such network disturbances is
treated in section 4.2. At this point it is assumed that the common-mode signal is a
harmonic alternating voltage whose amplitude may be a few volts and, in the case of
the mains disturbance, has the mains frequency (50 Hz or 60 Hz). The common-mode
signal also enters the measurement electronics. Figure 4.6 clarifies the measurement
situation. The amplitude of the common-mode signal (interference signal) often ex-
ceeds that of the biosignal by several orders of magnitude. From this, the necessity of
a very high common-mode rejection is directly derived as an important requirement
for the amplifier (Common Mode Rejection, in short: CMR). Eliminating the common-
mode signal with analog filters is not an option for most applications because the fil-
ter impedance is small and the rejection frequency band of the filter overlaps the fre-
quency band of the biosignal and would lead to distortion of the biosignal. This aspect
is discussed in detail in section 4.4.